AED
| Device class | Class II B under 93/42/EEC |
| Design standards | IEC 60601-1:2005+A1:2012, EN 60601-1-2:2015, IEC 60601-2-4:2010, ERC, ILCOR 2015 |
| Operating temperature | 0°C - 50°C |
| Storage temperature | -30°C - +70°C |
| Humidity | 10% - 95% relative humidity |
| Ingress protection | IP-55 |
| Operating mode | Semi-automatic / fully automatic |
| Waveform | Truncated biphasic waveform |
| Energy steps | Adult: 150J - 200J - 200J / Pediatric: 50J - 50J - 50J |
| Fast shock | < 10 seconds |
| Charge time | < 10 seconds |
| ECG analysis time | < 10 seconds |
| Data recording | Up to 12 hours ECG signal, event and ambient audio recording |
| CPR support | Available |
| Pacemaker detection | Available |
| Motion detection algorithm | Available |
| Warranty | 5 years |